Cerner Order Name IgE Immunoglobulin E (IgE), Blood
Performing Laboratory
St. Louis Children's Hospital
Lab Dept
Core Lab Chemistry
Specimen Requirements
Specimen Type: Blood
Container/Tube
Preferred: Light green-top (lithium heparin) Vacutainer®
Acceptable: green SAMPLETTE™/MICROTAINER®, plain red-top vacutainer ® or Gold/Red SST gel vacutainer®
Collection Volume: 1.0 mL (minimum 0.6 mL)
Patient Preparation: None
Sample Rejection: Mislabeled or unlabeled specimens.
Special Instructions: Send to lab promptly.
Logistics
Test Availability: Monday-Sunday
Cut-off time: None; performed as received
Turnaround Time: Same day; 2 hours from receipt in lab
Lab Testing Section: Chemistry
Interpretive
Reference Range:
< 1 year: <50.0 IUnits/mL
1 year - 3 years: <100.0 IUnits/mL
3 years - 5 years: <300.0 IUnits/mL
5 years - 8 years: <400.0 IUnits/mL
8 years - 16 years: <600.0 IUnits/mL
16 years - 18 years: <500.0 IUnits/mL
≥ 18 years: <100.0 IUnits/ML
Critical Values: No critical values established
Limitations: In rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies (such as HAMA), streptavidin or ruthenuim can occur.
Methodology: Electrochemiluminescence immunoassay (ECLIA) sandwich principle
Additional Information
For SLCH Laboratory use only
Lab Processing instructions: Centrifuge specimen to separate. If specimen is collected in microtainer or plain, red-top tube aliquot sample into a conical tube for testing (labeling aliquot with specimen type).
Processed Volume: minimum 0.3 mL plasma/serum
Analyte Stability:
Ambient - unprocessed: 8 hours
PROCESSED
Ambient: 8 hours
Refrigerated: 7 days
Frozen: 6 months
Notes: Will also accept Lavender-top (EDTA) Vacutainer®
General Information
EPIC order name: IgE (EPIC Test# LAB74)
Cerner order name: IgE
CPT Codes: 82785
LOINC Value: 19133-0
Synonyms:
Last Reviewed
November 2024